Nitrate (NO2-) Ion Selective Electrode
-
Specifications:
- DC resistance at 25°C ≤ 5.0 MOhm
- Minimum sample volume = 5 ml
- Preconditioning / Standard solution : NaNO3 1000 ppm as (NO3)
- Preconditioning time : at least 5 minutes
- Optimum pH range : pH 4.6 to pH 8
- Temperature range : 5°C to 50°C
- Optimum temperature : 25°C
- Recommended ISAB : 2M (NH4)2SO4 or 1M KAl(SO4)2 (Stock concentration to be diluted)
- Recommended reference electrode : Double Junction Reference electrode outer filling solution
:0.04M (NH4)2SO4 or 0.1M CH3COOLi or 0.1M KAl(SO4)2 - Electrode slope at 25°C 54 ± 5 mV/decade
- Linear measuring range : 3.1 to 62,000 ppm
- Total measuring range : 0.4 to 62,000 ppm
- Detection limit : 0.4 ppm
- Response time (time to complete 90% of the change in potential after immersion in the new
solution.) < 10 seconds - Time for stable reading after immersion 2 to 3 minutes
- Potential drift (in 1000 ppm) < 3 mV/day
- Selectivity coefficients k(NO3-: X)
Cl- : 5.6 x 10-3
HCO3- : 5 x 10 -3
NO2- : 1 x 10 -3
CH3COO- : 4.5 x 10 -4
F- : 1 x 10-4- Recalibration Frequency : Depending on precision requirements -