• Specifications:

    • DC resistance at 25°C ≤ 5.0 MOhm
    • Minimum sample volume = 5 ml
    • Preconditioning / Standard solution : NaNO3 1000 ppm as (NO3)
    • Preconditioning time : at least 5 minutes
    • Optimum pH range : pH 4.6 to pH 8
    • Temperature range : 5°C to 50°C
    • Optimum temperature : 25°C
    • Recommended ISAB : 2M (NH4)2SO4 or 1M KAl(SO4)2 (Stock concentration to be diluted)
    • Recommended reference electrode : Double Junction Reference electrode outer filling solution
      :0.04M (NH4)2SO4 or 0.1M CH3COOLi or 0.1M KAl(SO4)2
    • Electrode slope at 25°C 54 ± 5 mV/decade
    • Linear measuring range : 3.1 to 62,000 ppm
    • Total measuring range : 0.4 to 62,000 ppm
    • Detection limit : 0.4 ppm
    • Response time (time to complete 90% of the change in potential after immersion in the new
      solution.)
      < 10 seconds
    • Time for stable reading after immersion 2 to 3 minutes
    • Potential drift (in 1000 ppm) < 3 mV/day
    • Selectivity coefficients k(NO3-: X)

    Cl- : 5.6 x 10-3
    HCO3- : 5 x 10 -3
    NO2- : 1 x 10 -3
    CH3COO- : 4.5 x 10 -4
    F- : 1 x 10-4

    • Recalibration Frequency : Depending on precision requirements -
1 of 2
  • Ammonium Determination in Water

    Equipment and Chemicals: Ammonium Ion selective electrode, 3051 Combination (recommended) or 1051 Half Cell Reference electrode (not needed for 3051): double junction with NaCl or MgSO4. Standard solution: 1000 ppm...

    Ammonium Determination in Water

    Equipment and Chemicals: Ammonium Ion selective electrode, 3051 Combination (recommended) or 1051 Half Cell Reference electrode (not needed for 3051): double junction with NaCl or MgSO4. Standard solution: 1000 ppm...

1 of 3