Ammonium (NH4+) Ion Selective Electrode
-
Specifications:
- Ion Selective Electrode: Ammonium (NH4+)
- Sensor Type: PVC "dry" contact
- Limits (ppm): 9,000 to 0.9
- Concentration Range (molar): 0.5 to 5x10^-5
- pH Range: 0 to 8.5
- Main Interferences: K+
- Standard Solution: 1000 ppm NH4+ (Cat. #: 051/1000s)
- Ionic Strength Adjusting Buffer: 5 M NaCl (Cat. #: NaCl/ISA)
- Reference Electrode: Double-Junction
- Filling Solution for Reference Electrodes: 1 M NaCl (Cat. #: NaCl/DJ)
Applications Notes
-
Ammonium Determination in Water
Equipment and Chemicals: Ammonium Ion selective electrode, 3051 Combination (recommended) or 1051 Half Cell Reference electrode (not needed for 3051): double junction with NaCl or MgSO4. Standard solution: 1000 ppm...
Ammonium Determination in Water
Equipment and Chemicals: Ammonium Ion selective electrode, 3051 Combination (recommended) or 1051 Half Cell Reference electrode (not needed for 3051): double junction with NaCl or MgSO4. Standard solution: 1000 ppm...