To eliminate sample matrix effects which render direct potentiometry inaccurate, a known addition method is adopted to analyze the Calcium content of fruit juices (fresh).
Equipment:
- Calcium Combination ion selective electrodes 3041
- pH/Ion Analyzer (recommended) or pH Meter with millivolt scale
- Accurate balance weighing to ± 0.0005g
- Magnetic stirrer
Reagents:
- 1000 ppm Ca2+ standard solution
- 0.1M KNO3 (ISAB)
- 0.1M NaOH
Standard Preparation:
To 36.68g of CaCI22H2O in a 1 liter volumetric flask add deionised water to the mark. This is 1000 ppm Ca2+ solution.
Reagent Preparation:
Dissolve 4g NaOH in deionised water in a 1 liter volumetric flask. Dilute to the mark. This is 0.1M NaOH.
ISAB filling solution: to 10.1g KNO3 in a 1 liter volumetric flask add deionised water to the mark. This is 0.1M KNO3.
Method:
- To a 250 ml beaker on a magnetic stirrer add 20 ml of fruit juice.
- To the sample add 20 ml 0.1M KNO3 (ISAB).
- Add 40 ml deionised water.
- Adjust the pH of the sample solution to 6.0 by the addition of 0.1M NaOH.
- Dilute to 100ml using deionised water by transferring the beakeri s content to a 100ml volumetric flask and diluting to the mark.
- Empty the contents of the flask into the 250 ml beaker and immerse the electrodes in the solution.
- Record the stable mV potential (E1).
- Add 10 ml of the 1000 ppm Ca2+ standard (ensure continuous stirring) and record the new stable mV potential (E2).
Note:
Using Ion Analyzer , this determination may be carried out directly in concentration units using the Activity (Concentration) mode.
Calculation:
Calculate E by E2 - E1 (mV).
The sample solution concentration is given by this equation.
Cu = unknown solution concentration
Cs = standard concentration
Vs = volume of standard
Vu = volume of unknown
DE = change in potential (mV)
S = slope of electrode (mV)
To get the actual concentration of Ca2+ in the fruit juice, multiply the calculated value by 5.